Energy Matching Biochemical Filters

PRODUCT CATEGORIES
INOUIRY
Contact Form Demo (#3)

The energy matching type biochemical filter is a customized narrowband bandpass filter specifically designed for automatic biochemical analyzers and multi-channel enzyme-linked immunosorbent assay (ELISA) analyzers. It is classified as type III biochemical filter in the industry

The biochemical analyzer comes standard with halogen tungsten lamp light source, which has a naturally uneven spectrum: 340nm ultraviolet energy is extremely low, and 546/578 visible light energy is extremely strong; Paired with silicon optoelectronic PD detectors, the difference in raw signals between different wavelength channels can reach several to ten times.
Ordinary uniform transmittance filters will exhibit:
Weak 340nm channel signal, baseline drift, and poor repeatability;
578/630nm signal saturation, AD sampling overflow;
The gain level difference of multi-channel circuits is large, and the calibration of the whole machine is complex.

After special design, the coating film system is customized in layers, and the peak transmittance of each filter is artificially differentiated to offset the “difference in spectral energy of the light source+difference in detector response”. Finally, the light energy output of all wavelength channels reaching the photoelectric tube tends to be equal, and all channels of the whole machine can share the same amplification gain, greatly simplifying the circuit and calibration process.

The energy matching type biochemical filter

Standard wavelength+energy matching transmittance specification

340nm=JGS1 fused silica; 380~700nm=B270
Unified bandwidth: FWHM=8~10nm; Size Φ 8/Φ 10 × 2.0mm; AOI ≤ 3 ° collimated optical path

Menter wavelength Model Customized peak transmittance Light source energy compensation Clinical project
340nm BP340-M ≥70% Halogen UV energy is the weakest, increasing the transmittance compensation ALT/AST/LDH/NADH rate method
380nm BP380-M ≥62% Weak UV region, medium to high transmittance Urea, reference calibration
405nm BP405-M ≥58% Blue light energy is moderate GGT、ALP
492nm BP492-M ≥52% Visible light energy increases, moderately attenuates Blood lipids and immune turbidity
510nm BP510-M ≥50% Green light has sufficient energy and is transmitted at a low pressure Creatinine, iron and magnesium
546nm BP546-M ≥42% The strongest band of green light, greatly attenuated to prevent saturation Bilirubin, urea nitrogen, calcium
578nm BP578-M ≥40% Yellow green light energy peak, lowest level transmitted Total protein, albumin, uric acid
600nm BP600-M ≥43% Red light energy drops, slightly rebounds inorganic phosphorus
630nm BP630-M ≥46% Red light has moderate energy Reference wavelength and turbidity compensation