Biochemical filters are specifically used in biochemical analyzers and enzyme-linked immunosorbent assay (ELISA) analyzers, and are the core components of optical systems. After six generations of product development, Borisun has formed three major series, with over 60 wavelengths and more than 200 models and specifications of products. The wavelength range covers ultraviolet, visible, and near-infrared, with a large number of optional wavelengths. There is a standard product available every 3nm-5nm to meet diverse needs.

 Borisun biochemical filters are designed according to the optical path system of different instruments, and are divided into standard biochemical filters, energy matching biochemical filters, and multi-channel spectral biochemical filters. The unique deep cutoff, rectangular waveform, and ultra-high signal-to-noise ratio design ensure that the product achieves good absorbance testing accuracy, linearity, and consistency of detection results. The product performance has reached an advanced level, and the maximum absorbance test can reach 4ABS or above.

We use plasma assisted deposition (PIAD) and ion beam assisted deposition (IAD) processes to manufacture these filters, ensuring stable spectral performance and long service life of the filters.

Mainstream standard wavelength

Center wavelength Model abbreviation Wavelength tolerance Min peak transmittance Testing Items
340nm BP340 ±1 nm ≥55% ALT、AST、LDH、CK、 Glucose (hexokinase method), NADH rate method [core UV channel]
380nm BP380 ±1 nm ≥65% Partial urea, specific protein, partial dual wavelength reference
405nm BP405 ±2 nm ≥70% GGT、ALP、 Nitrophenol substrate reaction
492nm BP492 ±2 nm ≥75% Blood lipids, enzyme-linked immunosorbent assay, and partial immune turbidity
510nm BP510 ±2 nm ≥75% Creatinine, iron, magnesium
546nm BP546 ±2 nm ≥78% Total bilirubin, direct bilirubin, urea nitrogen, calcium
578nm BP578 ±2 nm ≥78% Total bilirubin, direct bilirubin, urea nitrogen, calcium
600nm BP600 ±2 nm ≥78% Phosphorus, partial immune transmittance turbidity
630nm BP630 ±2 nm ≥78% GLU、 Hemoglobin, reference wavelength, turbidity compensation
narrow band filters

Expand alternative wavelengths

BP410、BP415、BP450、BP505、BP550、BP565、BP660、BP670、BP700、BP750、BP800nm

BP505: Uric acid and triglycerides commonly used

BP700: mostly used as a dual wavelength reference channel to eliminate sample turbidity interference

Distinguishing between two types of equipment selection

1. Large scale fully automatic turntable biochemical analyzer (compatible with Hitachi/Olympus systems)
Standard configuration: 340, 405, 510, 546, 578, 600, 630
Unified bandwidth of 10nm, thickness of 2.0mm, mostly with a diameter of 10mm
2. Semi automatic biochemical/enzyme-linked immunosorbent assay (ELISA) analyzer
Commonly used: 340, 405, 450, 492, 505, 546, 630;
Some enzyme-linked immunosorbent assay (ELISA) systems will use the 8nm narrow bandwidth version to enhance their resistance to stray light